Jn. Verney et Am. Morgan, Evaluation of in vivo dose measurements for patients undergoing electron boost treatments, RADIOTH ONC, 59(3), 2001, pp. 293-296
Citations number
12
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Onconogenesis & Cancer Research
This study evaluated p-type silicon diodes for use in in vivo dosimetry in
clinical electron beams. A calibrated p-type silicon diode detector was use
d to measure the dose received by the patient in the centre of the held. Re
adings were corrected for energy, temperature and stand-off of the electron
applicator from the patient surface. The mean difference between measured
and prescribed dose was 1.04% (95% CI 0.72 to 1.36 %). (C) 2001 Elsevier Sc
ience Ireland Ltd. All rights reserved.