P. Westbroek et E. Temmerman, In line measurement of chemical oxygen demand by means of multipulse amperometry at a rotating Pt ring - Pt/PbO2 disc electrode, ANALYT CHIM, 437(1), 2001, pp. 95-105
In this paper an amperometric method that makes use of a the rotating ring-
disc electrode (RRDE) configuration is described for the determination of t
he chemical oxygen demand (COD). It is shown that the strong oxidant necess
ary in COD measurements can be generated by an electrolytic in situ formati
on of an aggressive species, at the disc part of the ring-disc electrode, t
hat oxidises the compounds that contribute to the GOD. The excess of aggres
sive species formed reacts further to oxygen, which is determined at the ri
ng electrode. Compensation of the disc current signal for the ring current
signal finally results into the COD value of the solution. In this paper th
e COD value of real samples from a textile and a photo developing company w
ere analysed and it was shown that a precision of 1% could be obtained with
in the range of 20-25,000 mg l(-1). An important advantage is the fact that
the COD value of a solution of unknown composition can be obtained within
1 min after calibration of the electrode system. (C) 2001 Elsevier Science
B.V. All rights reserved.