The main concern of producers of certified reference materials (CRM) is the
preparation of high-quality products with demonstrated homogeneity and sta
bility, combined with a well established set of certified characteristics.
CRM producers must, furthermore, comply with other constraints imposed by t
he ISO Guide 34:
production processes, production control, and certification analyses should
be performed by expert laboratories, using validated protocols documented
in their respective quality assurance manuals;
laboratory mean values and the corresponding "expanded" uncertainties, must
be used for the determination of the certified values, as recommended by t
he ISO Guide to the Expression of Uncertainties in Measurements (GUM); and
when possible, traceability of the certified value to the SI units, using a
ppropriately validated and/or primary methods, must be ensured.
k(0)-NAA, i.e. neutron activation analysis with k(0) standardization, is on
e of the analytical techniques implemented at the Reference Material Unit o
f IRMM; it meets the first two requirements.