The need for selenium speciation in an unknown bacterial sample brought by
a customer initiated the setup of a hyphenated method of IEC-ICP-MS, First
a published method from Gilon ct al. (J. Chromatogr. A. 1996, 750, 327) was
modified to analyze selenurea, seleno ethionine, seleno methionine (SeM),
Se(lv), seleno cystine (SeC) and Se(VI) by a hyphenated ICP-MS technique. L
ODs of 4-15 mug L-1 were achieved and a 1/10 diluted protease digestion (Pr
onase E) of the Se-containing bacterial sample was measured. However. the r
uggedness of the method was insufficient. Separation was compromised even a
t a 1/10 dilution of the sample and identification of the species was poor.
Another method was set up and developed to analyze anionic Se species. Thi
s method used a Dionex AS 11 column with NaOH and tetramethylammonium hydro
xide as eluents. Now, LODs of ca. 0.1 mug Se L-1 were achieved and the meth
od proved to be very rugged. No alteration of the separation was observed.
During 30 consecutive runs, retention times and peak areas varied only slig
htly in spite of the high material load on the column (e.g., for the major
compound the retention time was 6.26 min +/- 6 s, and the peak area variati
on was +/-0.7%). The protease digestion of the bacterial sample could be an
alyzed without dilution. The separation was not compromised, although the b
acterial material load of the sample was 30 mg mL(-1). Therefore low concen
tration Se species were not below the detection limit. Several species were
identified, such as Se(lv), SeC and in small amounts SeM. However, some sp
ecies remained unidentified due to unavailable standards For comparison. Ma
ss balances were also performed. The protease digestion provided 25% Se of
the total Se determined after pressure digestion (148 mg kg(-1)), and 98.7%
25% recovery was obtained when the protease digestion was speciated.