Wd. James et Mr. Raulerson, Evaluation of digital pulse processing-based gamma-ray spectroscopy under neutron activation analysis conditions, J RAD NUCL, 248(2), 2001, pp. 283-287
Digital signal processors are now available commercially for incorporation
into high resolution gamma-ray spectroscopy systems. In this work, we have
compared throughput peak resolution and peak stability found in two Canberr
a 2060 Digital Spectrum Processors with a conventional analog processing se
tup in our laboratory. We have made the comparisons for five separate high
purity germanium detectors which provide a range in detector size and const
ruction. In addition, the range of input count rates chosen for study refle
ct those likely to be encountered in NAA Our initial results indicate the p
erformance to be detector specific and highly dependent on DSP setup parame
ters.