Evaluation of digital pulse processing-based gamma-ray spectroscopy under neutron activation analysis conditions

Citation
Wd. James et Mr. Raulerson, Evaluation of digital pulse processing-based gamma-ray spectroscopy under neutron activation analysis conditions, J RAD NUCL, 248(2), 2001, pp. 283-287
Citations number
4
Categorie Soggetti
Inorganic & Nuclear Chemistry
Journal title
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY
ISSN journal
02365731 → ACNP
Volume
248
Issue
2
Year of publication
2001
Pages
283 - 287
Database
ISI
SICI code
0236-5731(200105)248:2<283:EODPPG>2.0.ZU;2-2
Abstract
Digital signal processors are now available commercially for incorporation into high resolution gamma-ray spectroscopy systems. In this work, we have compared throughput peak resolution and peak stability found in two Canberr a 2060 Digital Spectrum Processors with a conventional analog processing se tup in our laboratory. We have made the comparisons for five separate high purity germanium detectors which provide a range in detector size and const ruction. In addition, the range of input count rates chosen for study refle ct those likely to be encountered in NAA Our initial results indicate the p erformance to be detector specific and highly dependent on DSP setup parame ters.