Hp. Chen et al., Biopolymer sequencing using a triple quadrupole mass spectrometer in the ESI nozzle-skimmer/precursor ion MS/MS mode, J AM SOC M, 12(7), 2001, pp. 846-852
Citations number
27
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY
A variety of model biopolymers, including oligonucleotides, oligosaccharide
s and a synthetic pharmaceutical agent were sequenced using a triple quadru
pole mass spectrometer equipped with an electrospray source and operated in
a scan mode referred to as pseudo-MS3. This scan mode consists of three st
eps: (1) in-source collision-induced dissociation (CID) in the nozzle-skimm
er (NS) region, (2) scanning of the fragment ions into the collision cell f
or further CID, and (3) passing of the secondary fragment ions through the
final mass filter at a preselected mass, generally corresponding to the mas
s of a terminal sequence ion for the biopolymer. The mass spectra are recor
ded. in the precursor ion MS/MS mode where ion selection and detection occu
r at the third stage of the triple quadrupole but the scan function is dete
rmined by the first stage. The advantages and Limitations in using this pse
udo-MS3 NS/precursor ion MS/MS scan mode for biopolymer sequencing are disc
ussed. (C) 2001 American Society for Mass Spectrometry.