A pixel read-out chip for the PILATUS project

Citation
C. Bronnimann et al., A pixel read-out chip for the PILATUS project, NUCL INST A, 465(1), 2001, pp. 235-239
Citations number
4
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
465
Issue
1
Year of publication
2001
Pages
235 - 239
Database
ISI
SICI code
0168-9002(20010601)465:1<235:APRCFT>2.0.ZU;2-D
Abstract
For the protein crystallography beamlime of the Swiss Light Source (SLS), w hich will be operational in summer 2001, the PILATUS detector (PIxeL Appara TUs for the SLS) is currently being built. The goal of the PILATUS project is a hybrid pixel system with a size of 40 x 40 cm(2) having 2000 x 2000 pi xels. The detector will be operated in single photon counting mode for 12 k eV X-rays. The prototype pixel read-out chip has a noise of 75 e(-) with se nsor, a low threshold dispersion and a maximum count rate of 1 MHz/pixel. D ue to the expected radiation dose in operation at the beamline, the chip is designed in radiation hard DMILL technology. (C) 2001 Elsevier Science B.V . All rights reserved.