Pm. Dohmen et al., A reliable test to detect impending pump failure - during long-term support on the Novacor (R) N100 Left Ventricular Assist System, TEX HEART I, 28(2), 2001, pp. 139-141
The Durastudy is a new surveillance protocol to detect impending pump failu
re in Novacor((R)) N100 left Ventricular Assist Systems implanted for the l
ong term. Our patient, a 54-year-old man with a history of dilated cardiomy
opathy and contraindications for heart transplantation, received a Novacor
pump in May 1995 and did not experience sustainable ventricular recovery du
ring the subsequent 3 years. After more than 3 years of support, symptoms o
f pump wear were detected in this patient, through application of the Duras
tudy protocol. This enabled us to electively exchange the pump at 3.8 years
. Our patient remained in New York Heart Association functional class 1 unt
il he died in July 1999 of causes unrelated to the pump, after some 1,514 d
ays of support. This, we believe, still constitutes a world record.