Rapid determination of albumin of urine in Delta Iota measurement mode with portable spectrophotometer of high sensitivity

Citation
Hw. Chen et al., Rapid determination of albumin of urine in Delta Iota measurement mode with portable spectrophotometer of high sensitivity, CHEM J CH U, 22(6), 2001, pp. 928-930
Citations number
12
Categorie Soggetti
Chemistry
Journal title
CHEMICAL JOURNAL OF CHINESE UNIVERSITIES-CHINESE
ISSN journal
02510790 → ACNP
Volume
22
Issue
6
Year of publication
2001
Pages
928 - 930
Database
ISI
SICI code
0251-0790(200106)22:6<928:RDOAOU>2.0.ZU;2-J
Abstract
A portable spectrophotometer developed by Changchun R & D Center for Analyt ical Instruments, which combines DeltaI measurement mode and a measurement mode together with a high sensitivity, good accuracy and low cost , is empl oyed to replace the traditional spectrophotometer and the sensitivity is en hanced. Therefore, a new rapid method with low cost and simplicity in opera tion for protein determination is developed. For this method, the linear ra nge is shown to be 0-0, 74 mug/mL , the detection limit is 0.01 mug/mL, and the analysis can be completed within 5 min. The albumin in urine is determ ined and all the results are confirmed by the standard method. Because of t he rapid analysis speed, less expenditure, more convenience and simplicity in operation of this method, it is possible to use this method to diagnose glycosuric nephropathy at an early stage, even do it at home.