T. Washizuka et al., Inappropriate discharges from an intravenous implantable cardioverter defibrillator due to T-wave oversensing, JPN CIRC J, 65(7), 2001, pp. 685-687
Citations number
8
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
This report describes the clinical management of 2 patients with ventricula
r fibrillation (VF) who received inappropriate shocks from an implantable c
ardioverter defibrillator (ICD) due to T-wave oversensing. Cardiac sarcoido
sis was confirmed as the underlying heart disease in 1 patient and idiopath
ic dilated cardiomyopathy in the other. Within 2 months after ICD implantat
ion, both patients received several inappropriate shocks during sinus rhyth
m. Stored electrograms showed decreased R-wave amplitudes and increased T-w
ave amplitudes. The ICD sensed both R- and T-waves as ventricular activatio
n, which met the rate criteria for VF treatment. Reprogramming the sensing
threshold in association with administration of a drug to slow the heart ra
te decreased the incidence of the inappropriate shocks in both patients, bu
t these palliative measures did not completely suppress the inappropriate s
hocks. To avoid T-wave oversensing, the repositioning or adding of a sensin
g lead is required. The potential risk of T-wave oversensing in ICD patient
s who have small R-wave amplitudes should be recognized.