Sensitive determination of pethidine in body fluids by surface ionization organic mass spectrometry

Citation
A. Ishii et al., Sensitive determination of pethidine in body fluids by surface ionization organic mass spectrometry, J CHROMAT B, 758(1), 2001, pp. 117-121
Citations number
17
Categorie Soggetti
Chemistry & Analysis
Journal title
JOURNAL OF CHROMATOGRAPHY B
ISSN journal
13872273 → ACNP
Volume
758
Issue
1
Year of publication
2001
Pages
117 - 121
Database
ISI
SICI code
1387-2273(20010705)758:1<117:SDOPIB>2.0.ZU;2-F
Abstract
We have presented a simple and sensitive method for determining pethidine, a narcotic analgesic drug in body fluids by gas chromatography (GC)/surface ionization organic mass spectrometry (SIOMS). Good linearity was obtained in the range of 0.625-25 ng/ml of whole blood and urine by mass chromatogra phy, and in the range of 0.05-2 ng/ml of whole blood by selected ion monito ring (SIM). Pethidine and diphenylpyraline (internal standard) were extract ed from body fluids with Bond Elut Certify cartridges: their recoveries wer e above 95%. The detection limits (signal-to-noise ratio=3) were estimated to be 0.2 ng/ml of whole blood or urine by mass chromatography, 0.02 ng/ml of whole blood by SIM. (C) 2001 Elsevier Science B.V. All rights reserved.