UTILIZATION OF STANDARDS GENERATED IN THE PROCESS OF THERMAL-DECOMPOSITION CHEMICALLY-MODIFIED SILICA-GEL FOR A SINGLE-POINT CALIBRATION OFA GC FID SYSTEM/
M. Prokopowicz et al., UTILIZATION OF STANDARDS GENERATED IN THE PROCESS OF THERMAL-DECOMPOSITION CHEMICALLY-MODIFIED SILICA-GEL FOR A SINGLE-POINT CALIBRATION OFA GC FID SYSTEM/, Talanta, 44(9), 1997, pp. 1551-1561
The utilization of a multicomponent gaseous standard mixture, containi
ng CO and CO2 and obtained by thermal decomposition of a so-called imm
obilized compound, for a single point calibration of a GC/FID system a
re described. The generation of such a mixture takes place as a result
of thermal decomposition of a sample of chemically modified silica ge
l placed in a heated chamber of thermal desorber coupled with the devi
ce being calibrated via a catalytic methanizer. The mean amount of the
analyte librated from unit mass of the gel was 0.71 mg g(-1) (RSD = 3
%) for carbon monoxide and 0.86 mg g(-1) (RSD = 3%) for carbon dioxide
. (C) 1997 Elsevier Science B.V.