Reproducibility conditions for the pulsed X-ray emission in a vacuum sparkdischarge

Citation
N. Georgescu et al., Reproducibility conditions for the pulsed X-ray emission in a vacuum sparkdischarge, NUKLEONIKA, 46, 2001, pp. S17-S20
Citations number
6
Categorie Soggetti
Physics
Journal title
NUKLEONIKA
ISSN journal
00295922 → ACNP
Volume
46
Year of publication
2001
Supplement
1
Pages
S17 - S20
Database
ISI
SICI code
0029-5922(2001)46:<S17:RCFTPX>2.0.ZU;2-O
Abstract
In our laboratory, a 1 kJ vacuum spark device is developed as a pulsed X-ra y source for X-ray detector calibration and applications such as X-ray micr oscopy and microlithography. After our previous works for the X-ray emissio n maximization (up to around 10 J per pulse), this paper analyses the influ ence of four different triggering systems on the X-ray emission reproducibi lity. The main conclusion is that a high-power, very fast high voltage trig ger pulse is required. In order to fulfill these conditions, a pulse transf ormer and an air spark gap have been added to the initial triggering device (a magnetic pulse compression circuit). With the new trigger pulse (20 J/p ulse, 50 ns rise-time, 22 kV amplitude in an open circuit) a reasonable X-r ay emission reproducibility is obtained.