Ch. Fan et al., An unmediated hydrogen peroxide biosensor based on hemoglobin incorporatedin a montmorillonite membrane, ANALYST, 126(7), 2001, pp. 1086-1089
Hemoglobin was incorporated in a montmorillonite membrane. Electrochemical
and spectroscopic studies revealed that the electron transfer reactivity an
d peroxidase activity of the protein were both enhanced. Nevertheless, its
structure was still maintained native-like in the membrane. An unmediated h
ydrogen peroxide biosensor was accordingly prepared. The calibration plot o
f this H2O2 sensor was linear in the range of 1.0 x 10(-6)-6.0 x 10(-4) mol
L-1. The relative standard deviation was 3.1% for six successive determina
tions at a concentration of 1.0 x 10(-4) mol L-1. The detection limit was 6
.0 x 10(-7) mol L-1. Possible interferences in real sample analyses are dis
cussed.