Flow injection amperometry for the determination of iodate in iodized table salt

Citation
J. Jakmunee et K. Grudpan, Flow injection amperometry for the determination of iodate in iodized table salt, ANALYT CHIM, 438(1-2), 2001, pp. 299-304
Citations number
18
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICA CHIMICA ACTA
ISSN journal
00032670 → ACNP
Volume
438
Issue
1-2
Year of publication
2001
Pages
299 - 304
Database
ISI
SICI code
0003-2670(20010703)438:1-2<299:FIAFTD>2.0.ZU;2-S
Abstract
A very simple flow injection amperometric system for determination of iodat e has been developed. Cyclic voltammetry was used to study reactions of iod ate, iodide and iodine on a glassy carbon electrode. A single line manifold of the dow system consists of a pulse-free reagent propulsion using a simp le Mariotte bottle, a home-made injection valve, and a cross flow electroch emical cell with a potentiostat for control a potential of glassy carbon wo rking electrode at +200 mV versus Ag/AgCl electrode. Using a solution of 1. 0% w/v NaCl, 0.02% w/v KI and 0.1 M HCl as a reagent and a flow rate of 2 m l/min, a linear calibration up to 25 mg/l IO3- with a detection limit of 0. 5 mg/l IO3- was obtained. Relative standard deviation for 10 injections of 10 mg/l IO3- solution was 1.7%. A sample throughput of 35 injections per ho ur was achieved. The system was successfully applied for determination of i odate in iodized table salts, validating by the AOAC titrimetric method. (C ) 2001 Elsevier Science B.V. All rights reserved.