Comparison of SeroMP IgA with four other commercial assays for serodiagnosis of Mycoplasma pneumoniae pneumonia

Citation
T. Watkins-riedel et al., Comparison of SeroMP IgA with four other commercial assays for serodiagnosis of Mycoplasma pneumoniae pneumonia, DIAG MICR I, 40(1-2), 2001, pp. 21-25
Citations number
15
Categorie Soggetti
Clinical Immunolgy & Infectious Disease",Microbiology
Journal title
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE
ISSN journal
07328893 → ACNP
Volume
40
Issue
1-2
Year of publication
2001
Pages
21 - 25
Database
ISI
SICI code
0732-8893(200105/06)40:1-2<21:COSIWF>2.0.ZU;2-H
Abstract
Acute phase serum samples from 23 patients with Mycoplasma pneumoniae pneum onia were examined for specific antibodies with microparticle agglutination (MAG)-assay, complement fixation (CF)-test, IgM-specific ELISA, and Immuno card-based IgM-EIA. A novel 3-h IgA-ELISA (SeroMP-IgA) was tested to assess its diagnostic value. For MAG, CF, IgM-ELISA, card-based IgM-EIA, and IgA- ELISA positive results were obtained in 87%, 87%, 91%, 87%, and 100%, respe ctively. Overall concordance was 78%. Specificity testing of 46 healthy blo od donors revealed an optimal diagnostic cut-off range of 22-30 IgA-ELISA b inding units(BU)/mL resulting in 91%-100% specificity and 100%-96% sensitiv ity for serologic diagnosis of acute M, pneumoniae infection. (C) 2001 Else vier Science Inc. All rights reserved.