Determination of amlodipine in human plasma by high-performance liquid chromatography with fluorescence detection

Citation
S. Tatar et S. Atmaca, Determination of amlodipine in human plasma by high-performance liquid chromatography with fluorescence detection, J CHROMAT B, 758(2), 2001, pp. 305-310
Citations number
11
Categorie Soggetti
Chemistry & Analysis
Journal title
JOURNAL OF CHROMATOGRAPHY B
ISSN journal
13872273 → ACNP
Volume
758
Issue
2
Year of publication
2001
Pages
305 - 310
Database
ISI
SICI code
1387-2273(20010715)758:2<305:DOAIHP>2.0.ZU;2-R
Abstract
A sensitive and specific HPLC method has been developed for the assay of am lodipine in human plasma. The assay involves derivatization with 4-chloro-7 -nitrobenzofurazan (NBD-Cl), solid-phase extraction on a silica column and isocratic reversed-phase chromatography with fluorescence detection. Nortri ptyline hydrochloride was used as an internal standard. The assay was linea r over the concentration range of 0.25-18.00 ng/ml. Both of the within-day and day-to-day reproducibility and accuracy were less than 11.80% and 12.00 %, respectively. The plasma profile following a single administration of 10 mg amlodipine to a healthy volunteer was presented. (C) 2001 Elsevier Scie nce B.V. All rights reserved.