Validation of the AmpFlSTR (TM) profiler plus PCR amplification kit for use in forensic casework

Citation
We. Frank et al., Validation of the AmpFlSTR (TM) profiler plus PCR amplification kit for use in forensic casework, J FOREN SCI, 46(3), 2001, pp. 642-646
Citations number
21
Categorie Soggetti
Research/Laboratory Medicine & Medical Tecnology
Journal title
JOURNAL OF FORENSIC SCIENCES
ISSN journal
00221198 → ACNP
Volume
46
Issue
3
Year of publication
2001
Pages
642 - 646
Database
ISI
SICI code
0022-1198(200105)46:3<642:VOTA(P>2.0.ZU;2-H
Abstract
According to TWGDAM guideline 4.5 (1), prior to implementing a new DNA anal ysis procedure or an existing DNA analysis procedure developed by another l aboratory, the forensic laboratory must first demonstrate reliability of th e procedure inhouse. Seven phases were designed to validate the use of the AmpFlSTR Profiler Plus PCR Amplification Kit, as well as the PE Applied Bio systems 310 Genetic Analyzer. This report summarizes the results obtained f or each of the seven phases of the validation study which included the foll owing evaluations: polymer, reproducibility, sensitivity, stutter, precisio n, mixtures and nonprobative casework.