Re. Stanford et al., A novel, resorbable suture anchor: Pullout strength from the human cadavergreater tuberosity, J SHOUL ELB, 10(3), 2001, pp. 286-291
The pullout strength of a collagen bone anchor that creates interference fi
xation as the result of radial swelling on hydration was compared with a Mi
tek rotator cuff anchor after insertion into the greater tuberosity of huma
n cadaver humeri. Bones were fully hydrated at 37 degreesC. Stiffness, peak
load, and the mode of failure were recorded. Real and apparent bone densit
ies were measured Peak load for the collagen anchor at 15 minutes (127.0N /- 81.3N) was greater than at 2 minutes (60.5N +/- 38.5N) after insertion (
P < .05). At between 5 and 60 minutes after insertion, peak loads for the M
itek and the collagen anchors did not differ. After 30 minutes From inserti
on, the mode of failure of the collagen anchor changed from pullout with mi
nor body damage to pullout with major body damage. Peak load at pullout cor
related with bone density for the Mitek (P < .05, r = 0.516) but for the co
llagen bone anchor appeared unaffected by bone density.