Abstract
The Large Area Detector is a photon counting detector based on pixel readou t ASICs with 64 x 64 arrays of 150 mum square pixels bump bonded to a 300 m um thick Si detector. It is designed for X-ray diffraction studies and is c apable of up to 1 MHz count rate per pixel. A prototype has been tested at the synchrotron radiation source in Daresbury in a 19.5 keV X-ray beam. Thi s paper focuses on the spatial resolution and noise performance. The calcul ation of modulation transfer functions leads to analytical formulae allowin g the determination of the theoretical spatial resolution. (C) 2001 Elsevie r Science B.V. All rights reserved.