The analysis of the automated blood cell count is an essential tool in haem
atological diagnostics. However, in the case of the white blood cell differ
ential the microscopy method, although tedious, often serves as reference.
We evaluated the ABX Pentra 120 Retic haematology analyser in comparison to
the Coulter STKS haematology system and the microscopy method with respect
to accuracy, precision and reliability. We compared 308 samples (239 sampl
es from adults and 69 from children) including patients with oncological di
seases. The comparison of the white blood cell differential revealed strong
correlations between the results obtained with the ABX Pentra 120 Retic an
d the microscopy method, the Coulter STKS and the microscopy method and bot
h automated methods (values of paediatric samples in parentheses; neutrophi
ls: r(s)greater than or equal to0.933 (r(s)greater than or equal to0.951).
lymphocytes: r(s)greater than or equal to0.907 (r(s)greater than or equal t
o0.945), monocytes: r(s)greater than or equal to0.584 (r(s)greater than or
equal to0.459) and eosinophils: r(s)greater than or equal to0.963 (r(s)grea
ter than or equal to0.966)). The analytical performance of automatic analys
ers for the detection of the morphological "left shift" was determined for
all samples in comparison to the microscopical white blood cell differentia
l. The sensitivity, specificity and efficiency depended strongly an the cho
sen threshold levels and were different for both analysers. The sensitivity
for flagging a left shift increased with an increasing proportion of neutr
ophil bands, metamyelocytes, myelocytes and promyelocytes. Our study sugges
ts that the ABX Pentra 120 Retic haematology analyser, as well as the Coult
er STKS haematology system are useful tools for routine analysis in haemato
logy.