Counting efficiency of condensation particle counters at low-pressures with illustrative data from the upper troposphere

Citation
M. Hermann et A. Wiedensohler, Counting efficiency of condensation particle counters at low-pressures with illustrative data from the upper troposphere, J AEROS SCI, 32(8), 2001, pp. 975-991
Citations number
30
Categorie Soggetti
Chemical Engineering
Journal title
JOURNAL OF AEROSOL SCIENCE
ISSN journal
00218502 → ACNP
Volume
32
Issue
8
Year of publication
2001
Pages
975 - 991
Database
ISI
SICI code
0021-8502(200108)32:8<975:CEOCPC>2.0.ZU;2-U
Abstract
The low-pressure counting efficiency of three modified condensation particl e counters (CPCs, TSI Model 7610) was investigated. In this study, we descr ibe the modifications of the counters, present a "new" low-pressure calibra tion set-up, and show the resulting CPC counting efficiency curves as funct ion of operating pressure (160-1000 hPa) and temperature difference between the saturator and condenser block (10.5-26.0 degreesC). Furthermore. parti cle coincidence in the CPC optics was studied. The presented CPCs are able to measure particle number concentrations of up to 40,000 particles/cm(3) a t pressures down to 160 hPa. The lower threshold diameter of the CPCs can b e varied between 4 and 23 nm. (C) 2001 Elsevier Science Ltd. All rights res erved.