Wd. Ehmann et al., DETERMINATION OF OXYGEN, FLUORINE AND BORON BY 14-MEV INAA - A CASE OF MULTIPLE INTERFERENCES, Journal of radioanalytical and nuclear chemistry, 180(2), 1994, pp. 277-284
Oxygen and fluorine have been simultaneously determined by 14-MeV INAA
in samples containing boron. Both boron and fluorine can cause seriou
s interferences in the determination of oxygen. The fluorine and boron
interference corrections for oxygen determination have been determine
d to be 0.43 +/- 0.01 and 0.0832 +/- 0.0017 apparent g oxygen per g of
fluorine and boron, respectively, for our system. Boron can be determ
ined in the same sample by a second irradiation. Mutual interferences
have been evaluated and the procedure has been applied to NIST SRM and
several other compounds.