APPLICATION OF RAPID ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRIC METHODS TO THE DETERMINATION OF AG, AL, CD AND MN IN COCAINE AND HEROINSAMPLES

Citation
P. Bermejobarrera et al., APPLICATION OF RAPID ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRIC METHODS TO THE DETERMINATION OF AG, AL, CD AND MN IN COCAINE AND HEROINSAMPLES, Fresenius' journal of analytical chemistry, 358(7-8), 1997, pp. 844-847
Citations number
10
Categorie Soggetti
Chemistry Analytical
ISSN journal
09370633
Volume
358
Issue
7-8
Year of publication
1997
Pages
844 - 847
Database
ISI
SICI code
0937-0633(1997)358:7-8<844:AOREAS>2.0.ZU;2-T
Abstract
Rapid methods were developed for the direct determination of Ag, Ali C d and Mn in cocaine and heroin by ETAAS using programmes omitting the charring step. Sample pretreatment was simple: dissolution in ultrapur e water or in 35.0% (v/v) HNO3 for heroin or cocaine respectively. Opt imum drying temperatures were 250 degrees C for Ag, Al and Mn, and 300 degrees C for Cd. The run cycles were 35 and 37 s, for Ag and Al resp ectively, and 36 s for Cd and Mn. The best results were obtained with Pd. Mg(NO3)(2) and (NH4)(2)HPO4, as chemical modifiers, The limits of detection were 8.6, 55.9, 2.2 and 12.4 mu g kg(-1) for Ag, Al, Cd and Mn, respectively.