E. Moguiline et al., DEVELOPMENT OF SILICON DRIFT DETECTORS AND DIGITAL FILTERING ELECTRONICS FOR X-RAY SPECTROSCOPY, Journal de physique. IV, 7(C2), 1997, pp. 339-340
We report on the performances of energy resolved detectors and electro
nics developed for X-ray absorption spectroscopy studies in the fluore
scence excitation mode. Large active area multianode silicon drift det
ectors have been designed and evaluated. A typical energy resolution o
f 170 eV FWHM has been measured at 5.98 keV for a 6 mu s triangular sh
aping time. The data acquisition is based on multichannel digital shap
ing amplifiers developed at the ESRF. Excellent energy resolutions can
be achieved with reduced dead times in comparison with analog shapers
. The high level of integration and the full software control is compa
tible with the implementation of large detector arrays.