THE REPRODUCIBILITY OF COLOR DOPPLER DUPLEX SONOGRAPHY IN THE MEASUREMENT OF RENAL ARTERIAL BLOOD VELOCITY

Citation
N. Imuta et al., THE REPRODUCIBILITY OF COLOR DOPPLER DUPLEX SONOGRAPHY IN THE MEASUREMENT OF RENAL ARTERIAL BLOOD VELOCITY, Ultrasound in medicine & biology, 23(6), 1997, pp. 813-819
Citations number
28
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging",Acoustics
ISSN journal
03015629
Volume
23
Issue
6
Year of publication
1997
Pages
813 - 819
Database
ISI
SICI code
0301-5629(1997)23:6<813:TROCDD>2.0.ZU;2-4
Abstract
In this study, the reproducibility of color Doppler duplex sonography for repeated measurements of renal blood flow was evaluated in 14 heal thy subjects. We examined the reproducibility for different examiners and different time intervals between the examinations. Doppler frequen cy sonograms were analyzed with several parameters, and statistical ev aluation was performed by calculating both the correlation coefficient (r) and the coefficient of variation (CV). Peak systolic velocity (S) , early diastolic velocity (D1) and mean velocity (MV) showed good rep roducibility (r = 0.902-0.992, CV = 2.15-8.16%). On the other hand, en d-diastolic velocity (D2), acceleration time (AT) and acceleration ind ex (AI) showed poor reproducibility. We conclude that the reproducibil ity of this method is acceptable for repeated measurements of renal bl ood velocity, using suitable parameters S, D1 and MV. (C) 1997 World F ederation for Ultrasound in Medicine & Biology.