T. Katsu et al., POTENTIOMETRIC FLOW-INJECTION DETERMINATION OF SERUM BROMIDE IN PATIENTS WITH EPILEPSY, Journal of pharmaceutical and biomedical analysis, 15(12), 1997, pp. 1829-1832
A flow injection system was constructed using a bromide-selective elec
trode and used to determine serum bromide in patients with epilepsy. A
10-mu l serum sample was injected into a carrier stream flowing at 0.
12 mi min(-1). Potential changes and bromide concentrations were linea
rly related in the range 3-50 mM. The lower limit of detection for ser
um bromide was 1 mM and this electrode sensitivity spanned the entire
concentration range required for bromide therapy (9-24 mM). The result
s compared favourably with those obtained by colorimetry. (C) 1997 Els
evier Science B.V.