AUTOCAPTURE AS A DIAGNOSTIC-TOOL - DETECT ION OF A LEAD DISLODGMENT

Citation
A. Liebold et De. Birnbaum, AUTOCAPTURE AS A DIAGNOSTIC-TOOL - DETECT ION OF A LEAD DISLODGMENT, Zeitschrift fur Kardiologie, 86(7), 1997, pp. 521-523
Citations number
6
Categorie Soggetti
Cardiac & Cardiovascular System
Journal title
ISSN journal
03005860
Volume
86
Issue
7
Year of publication
1997
Pages
521 - 523
Database
ISI
SICI code
0300-5860(1997)86:7<521:AAAD-D>2.0.ZU;2-3
Abstract
Autocapture is a newly developed automatic threshold tracking algorith m implemented in implantable pacemakers. Based on the detection of the cardiac evoked response, autocapture continuously controls the effect iveness of stimulation. Increases in threshold are immediately followe d by an adaption of the pulse amplitude. Automatic threshold tests per iodically lower the amplitude in order to adapt to decreased threshold s. Thus, changes in threshold can be monitored over a long period and utilized as a diagnostic tool. This case report demonstrates the diagn osis and monitoring of a microdislodgement by means of autocapture.