K. Ekstrand et al., THE INFLUENCE OF X-RAY-ENERGY ON LUNG DOSE UNIFORMITY IN TOTAL-BODY IRRADIATION, International journal of radiation oncology, biology, physics, 38(5), 1997, pp. 1131-1136
Citations number
11
Categorie Soggetti
Oncology,"Radiology,Nuclear Medicine & Medical Imaging
Purpose: In this study we examine the influence of x-ray energy on the
uniformity of the dose within the lung in total-body irradiation trea
tments in which partial transmission blocks are used to control the lu
ng dose, Methods and Materials: A solid water phantom with a cork inse
rt to simulate a lung was irradiated by x-rays with energies of either
6, 10, or 18 MV, The source to phantom distance was 3.9 meters. The c
ork insert was either 10 cm wide or 6 cm wide, Partial transmission bl
ocks with transmission factors of 50 % were placed anterior to the cor
k insert, The blocks were either 8 or 4 cm in width, Kodak XV-2 film w
as placed in the midline of the phantom to record the dose, Midplane d
ose profiles were measured with a densitometer, Results: For the 10 cm
wide cork insert the uniformity of the dose over 80% of the block wid
th varied from 6.6% for the 6 MV x-rays to 12.2% for the 18 MV x-rays.
For the 6 cm wide cork insert the uniformity was comparable for all t
hree x-ray energies, but for 18 MV the central dose increased by 9.4%
compared to the 10 cm wide insert, Conclusion: Many factors must be co
nsidered in optimizing the dose for total-body irradiation, This study
suggests that for AP/PA techniques lung dose uniformity is superior w
ith 6 MV irradiation, The blanket recommendation that the highest x-ra
y energy be used in TBI is not valid for all situations. (C) 1997 Else
vier Science Inc.