CRANIOFACIAL OSSEOINTEGRATED IMPLANT-INDUCED STRAIN DISTRIBUTION - A NUMERICAL STUDY

Citation
V. Delvalle et al., CRANIOFACIAL OSSEOINTEGRATED IMPLANT-INDUCED STRAIN DISTRIBUTION - A NUMERICAL STUDY, The International journal of oral and maxillofacial implants, 12(2), 1997, pp. 200-210
Citations number
15
Categorie Soggetti
Dentistry,Oral Surgery & Medicine
ISSN journal
08822786
Volume
12
Issue
2
Year of publication
1997
Pages
200 - 210
Database
ISI
SICI code
0882-2786(1997)12:2<200:COISD->2.0.ZU;2-H
Abstract
Strain distributions that occur in the hard tissue in the region surro unding craniofacial osseointegrated implants are compared. Three comme rcially available implant designs were evaluated under bob axial loadi ng and axial loading with a moment in three bone configurations typica l of the craniofacial region. The evaluations that used the finite ele ment method indicated that for axial loading, the implant designs prod uced similar strain levels in each hone configuration. When moments as well as vertical loads were applied, the strains were three to seven times higher and variations among the designs were greater. The variat ions found were related to the amount of bone present in each situatio n, as well as the neck diameter of the implant involved.