X. Fei et al., PRECISION-MEASUREMENT OF THE MAGNETIC-FIELD IN TERMS OF THE FREE-PROTON NMR FREQUENCY, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 394(3), 1997, pp. 349-356
We have developed a NMR standard probe using a spherical pure water sa
mple of 1 cm diameter to determine the absolute magnetic field B in te
rms of the free-proton NMR frequency f(p) with an accuracy of 3.4 x 10
(-8). Our standard probe can be used conveniently to calibrate other p
robes in the field range from 1.45 to 1.7T and can readily be employed
over a much wider field range. The probe design and the tests carried