A. Belay et al., LC-BIOSENSOR SYSTEM FOR THE DETERMINATION OF THE NEUROTOXIN BETA-N-OXALYL-L-ALPHA,BETA-DIAMINOPROPIONIC ACID, Analytical chemistry, 69(17), 1997, pp. 3471-3475
An analysis system is described for the determination of the neurotoxi
n beta-N-oxalyl-alpha,beta-diaminopropionic acid (beta-ODAP). The syst
em is based on liquid chromatographic separation of beta-ODAP from int
erfering amino acids on an anion exchange column coupled with an amper
ometric enzyme electrode for the registration of beta-ODAP. The electr
ode is based on a graphite rod modified with an Os2+/3+ redox polymer
cross-linked with L-glutamate oxidase and horseradish peroxidase. This
LC-bienzyme electrode analytical system enabled monitoring of as low
as 4 mu M beta-ODAP (injection volume 100 mu L). The beta-ODAP content
in real grass pea samples was measured to range between 3.3 and 5.2 g
kg(-1) in dry grass pea.