DEVELOPMENT OF A DETECTOR FOR ULTRATRACE NITROGEN IN ARGON USING LOW-PRESSURE, CAPILLARY GLOW-DISCHARGE MOLECULAR-EMISSION SPECTROPHOTOMETRY
Citation
H. Ogino et T. Seki, DEVELOPMENT OF A DETECTOR FOR ULTRATRACE NITROGEN IN ARGON USING LOW-PRESSURE, CAPILLARY GLOW-DISCHARGE MOLECULAR-EMISSION SPECTROPHOTOMETRY, Analytical chemistry, 69(17), 1997, pp. 3636-3640
Categorie Soggetti
Chemistry Analytical
SICI code
0003-2700(1997)69:17<3636:DOADFU>2.0.ZU;2-Z
Abstract
A highly sensitive detector was developed for determining parts per bi
llion of nitrogen in high-purity argon. The method is based on the spe
ctrophotometric determination of the emission intensity from the nitro
gen molecule excited in the lowest vibrational transitions of the elec
tronic C-3 Pi(u) --> B-3 Pi(g) system. A capillary glow discharge tech
nique was applied to excite nitrogen in an argon now at several Torr.
The low-pressure, capillary glow discharge method offered high sensiti
vity for detecting trace nitrogen in argon. The response of the detect
or was proportional in the range of nitrogen concentrations from 0.1 p
pb, which correspond to S/N = 3, to 100 ppm.