Ml. Oldfield et al., MEASUREMENTS OF THE SIDE-BAND CONVERSION GAIN RATIO OF A MILLIMETER-WAVE HETERODYNE SUB-HARMONIC MIXER USING A FOURIER-TRANSFORM SPECTROMETER, International journal of infrared and millimeter waves, 18(8), 1997, pp. 1547-1563
A Fourier transform spectrometer, based on a Martin-Puplett polarisati
on rotation interferometer and using broad-band blackbody noise source
s, has been us-ed to study the sideband response i.e., conversion gain
, of a room temperature Schottky diode sub-harmonic mixer operating at
300 GHz. The technique enables the response of the mixer to be charac
terised and preferentially tuned to one sideband, thereby improving th
e rejection of unwanted spectral components which can be present in th
e mixer image sideband.