CT-GUIDED INTERVENTION BY ON-TARGET LASER GUIDANCE

Citation
R. Kloppel et al., CT-GUIDED INTERVENTION BY ON-TARGET LASER GUIDANCE, RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der neuenbildgebenden Verfahren, 167(2), 1997, pp. 194-197
Citations number
11
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
09366652
Volume
167
Issue
2
Year of publication
1997
Pages
194 - 197
Database
ISI
SICI code
0936-6652(1997)167:2<194:CIBOLG>2.0.ZU;2-H
Abstract
Purpose: The present study evaluates the use of a laser guidance syste m far CT-guided intervention, Material and methods: 94 cases of diagno stic biopsies and lumbar sympathectomies (54 cases with laser guidance system and 40 without) were compared. Results: Using the laser guidan ce system, the number of control scans decreased by 30 to 50%, and nec essary corrections of needle location were reduced by a maximum of 30% . The average target: deviation of the needle decreased to less than 5 mm in 50% of cases. Conclusion: The laser guidance system is strongly recommended in CT-guided interventions for quality assurance and high er efficiency. The advantage is especially marked if the target area i s small.