RETENTION TIME REPEATABILITY AS A FUNCTION OF THE INJECTION AUTOMATISM IN THE ANALYSIS OF TRACE ORGANOCHLORINATED COMPOUNDS WITH HIGH-RESOLUTION GAS-CHROMATOGRAPHY

Citation
L. Berdie et al., RETENTION TIME REPEATABILITY AS A FUNCTION OF THE INJECTION AUTOMATISM IN THE ANALYSIS OF TRACE ORGANOCHLORINATED COMPOUNDS WITH HIGH-RESOLUTION GAS-CHROMATOGRAPHY, Journal of chromatography, 778(1-2), 1997, pp. 23-29
Citations number
11
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Journal title
Volume
778
Issue
1-2
Year of publication
1997
Pages
23 - 29
Database
ISI
SICI code
Abstract
Short-term precision in retention time determination of different inje ction systems (three manual and three autosamplers) is assessed in the present study. This evaluation has been performed on eight different instrumental set-ups equipped with the same type of capillary column ( 30 m, 5% phenyl-95% methyl) which have been used for the analysis of s tandard mixtures of alpha-, gamma-, and delta-hexachlorocyclohexanes. The results show that, as expected, retention time repeatability incre ases with the degree of automatization of the injection system. They a lso show that only the autosamplers and the manual injector provided w ith the syringe-operated start device are suitable for analysis of com pounds with retention time differences in the order of 0.1 min. Conver sely, none of the systems included in the study, not even the autosamp lers, is adequate for the analysis of analytes having retention time d ifferences of about 0.01 min. (C) 1997 Elsevier Science B.V.