ANALYSIS OF VENTRICULAR WALL-MOTION USING COLOR-CODED TISSUE DOPPLER IMAGING-SYSTEM
Citation
N. Yamazaki et al., ANALYSIS OF VENTRICULAR WALL-MOTION USING COLOR-CODED TISSUE DOPPLER IMAGING-SYSTEM, JPN J A P 1, 33(5B), 1994, pp. 3141-3146
Categorie Soggetti
Physics, Applied
Abstract
We developed a new color Doppler system by which Doppler signals assoc
iated with tissue motion can be determined, and called it the tissue D
oppler imaging (TDI) system. Using high-speed scanning, the frame rate
was 26-38 F/s, and the pulse repetition frequency was 4.5-6.0 kHz. Un
der these conditions, the lowest measurable velocity was improved to 2
mm/s. Wall motion toward the transducer was coded as red, and that aw
ay from the transducer as blue. To examine the accuracy and validity o
f the measured velocity of ventricular wall motion, we performed in vi
tro and in vivo studies. The results demonstrate that the present TDI
system accurately in real time, facilitating the visual assessment of
abnormal ventricular wall motion.