MEASUREMENT OF SHOCK-WAVE PRESSURE DISTRIBUTION OF EXTRACORPOREAL SHOCK-WAVE LITHOTRIPTER USING PRESSURE-SENSITIVE PAPER - PRACTICAL USEFULNESS AND LIMITATIONS

Citation
M. Oyanagi et al., MEASUREMENT OF SHOCK-WAVE PRESSURE DISTRIBUTION OF EXTRACORPOREAL SHOCK-WAVE LITHOTRIPTER USING PRESSURE-SENSITIVE PAPER - PRACTICAL USEFULNESS AND LIMITATIONS, JPN J A P 1, 33(5B), 1994, pp. 3155-3158
Citations number
9
Categorie Soggetti
Physics, Applied
Volume
33
Issue
5B
Year of publication
1994
Pages
3155 - 3158
Database
ISI
SICI code
Abstract
A new, method for evaluating shock wave pressure distribution of extra corporeal shock wave lithotripters, using pressure-sensitive paper has been developed. It is a simple method to visualize the shock wave pre ssure distribution and has a better spatial resolution than the hydrop hone. The pressure-sensitive paper can be used to visually detect whet her the focal point is on the correct position, and to detect changes of pressure distribution. Additionally, the relationship between litho tripter power and integrated density, as a new evaluating value shows linearity. Therefore, this method is useful for quality assurance of l ithotripters at hospitals.