M. Nauck et al., MULTICENTER EVALUATION OF A HOMOGENEOUS ASSAY FOR HDL-CHOLESTEROL WITHOUT SAMPLE PRETREATMENT, Clinical chemistry, 43(9), 1997, pp. 1622-1629
We evaluated a new homogeneous assay for the measurement of HDL-choles
terol (HDL-C) in six European laboratories. The assay includes two rea
gents and is applicable to most autoanalyzers, which allows full autom
ation. The total CVs of the new method ranged between 1.3% and 6.7%. T
hereby determined HDL-C values were in good agreement with those obtai
ned by precipitation with phosphotungstic acid/MgCl2 or by a combinati
on of ultracentrifugation and precipitation (0.956< r <0.994). The ass
ay was linear up to at least 1500 mg/L HDL-C. Hemoglobin did not inter
fere, whereas icteric samples with bilirubin >100 mg/L showed discrepa
ncies between the homogeneous and the precipitation assay, Lipemia up
to total triglyceride concentrations of 8000 mg/L did not interfere wi
th the homogeneous HDL-C assay. The homogeneous HDL-C assay was easy t
o handle and produced similar results in all laboratories participatin
g in this study. This method will significantly facilitate the screeni
ng of individuals at increased risk for cardiovascular disease.