HOLMIUM-YAG LASER DAMAGE TO URETERAL GUIDEWIRE

Citation
Gs. Freiha et al., HOLMIUM-YAG LASER DAMAGE TO URETERAL GUIDEWIRE, Journal of endourology, 11(3), 1997, pp. 173-175
Citations number
14
Categorie Soggetti
Urology & Nephrology
Journal title
ISSN journal
08927790
Volume
11
Issue
3
Year of publication
1997
Pages
173 - 175
Database
ISI
SICI code
0892-7790(1997)11:3<173:HLDTUG>2.0.ZU;2-U
Abstract
The holmium:yttrium-aluminum-garnet (holmium:YAG) laser effectively fr agments urinary calculi of all compositions. Complications have been m inimal and mostly attributable to the ureteroscopy as opposed to the l aser. We report a case in which the polytetrafluoroethylene (PTFE) coa ting of a ureteral guidewire was broken with the holmium:YAG laser and remained in the collecting system after the guidewire was removed pos toperatively. The patient required cystoscopy to remove the guidewire coating.