AMPEROMETRIC DETERMINATION OF HYDROGEN-PEROXIDE WITH A MANGANESE DIOXIDE-MODIFIED CARBON-PASTE ELECTRODE USING FLOW-INJECTION ANALYSIS

Citation
K. Schachl et al., AMPEROMETRIC DETERMINATION OF HYDROGEN-PEROXIDE WITH A MANGANESE DIOXIDE-MODIFIED CARBON-PASTE ELECTRODE USING FLOW-INJECTION ANALYSIS, Analyst, 122(9), 1997, pp. 985-989
Citations number
35
Categorie Soggetti
Chemistry Analytical
Journal title
ISSN journal
00032654
Volume
122
Issue
9
Year of publication
1997
Pages
985 - 989
Database
ISI
SICI code
0003-2654(1997)122:9<985:ADOHWA>2.0.ZU;2-Z
Abstract
A carbon paste electrode bulk modified with MnO2 was investigated as a n amperometric detector for hydrogen peroxide in flow injection analys is (FIA). With an operating potential of +0.46 V versus Ag/AgCl, H2O2 produces catalytic oxidation currents which can be exploited for quant itative determinations. Factors influencing the analytical performance of the electrode, such as paste composition and pH, were studied both for batch voltammetry and for FIA. For the flow system the influence of the injection volume and flow rate were examined. The amperometric signals are linearly proportional to H2O2 concentrations in the range 0.5-350 mg l(-1), showing a detection limit (three times the signal-to -noise ratio) of 45 mu g l(-1).