NEW RESULTS WITH SILICON PAD DETECTORS AND LOW-NOISE ELECTRONICS FOR ABSORPTION SPECTROMETRY

Citation
A. Pullia et al., NEW RESULTS WITH SILICON PAD DETECTORS AND LOW-NOISE ELECTRONICS FOR ABSORPTION SPECTROMETRY, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 395(3), 1997, pp. 452-456
Citations number
8
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
395
Issue
3
Year of publication
1997
Pages
452 - 456
Database
ISI
SICI code
0168-9002(1997)395:3<452:NRWSPD>2.0.ZU;2-6
Abstract
High-resistivity silicon diode detectors integrable in array structure s have been tested for high-resolution X-ray spectroscopy at various t emperatures. Using an inexpensive commercially available front-end tra nsistor we have obtained resolutions of 19 electrons rms, or 165 eV FW HM, at -35 degrees C and 39 electrons rms, or 320 eV FWHM, at 26 degre es C, with 3 and 1 mu s shaping times. The obtained resolution values are adequate for synchrotron light based EXAFS experiments. These reli able detectors can be suitably employed for high-performance multichan nel systems in which high-rate capability is achieved by using an arra y of detectors operated in parallel. A 16-element detector prototype h as been recently assembled, installed and successfully operated for a real EXAFS experiment at NSLS.