The exit wave function in coherent STEM can be factorised in a probe f
unction and an object function for non-phase objects under certain res
trictions on specimen thickness and probe smoothness. This has importa
nt consequences for direct reconstruction of the specimen structure fr
om electron microscope images: e.g. for phase retrieval and deconvolut
ion techniques.