Ks. Wong et Es. Yeung, SIMULTANEOUS MONITORING OF GLUTATHIONE AND MAJOR PROTEINS IN SINGLE ERYTHROCYTES, Mikrochimica acta, 120(1-4), 1995, pp. 321-327
To gain more information from single-cell analysis, multiple detection
schemes can be implemented in capillary electrophoresis. Red blood ce
lls pre-treated with monobromobimane allow fluorescent tags to be atta
ched to various thiol-containing compounds. Both derivatized glutathio
ne and derivatized hemoglobin-A(0) can be detected by observing emissi
on at 495 nm when 275 nm excitation is used. At the same time, a secon
d emission channel at 350 nm can be used to monitor the native fluores
cence from carbonic anhydrase and from hemoglobin. Cross-correlation o
f the intracellular amounts of these provides additional insight into
the cell-aging process.