GLASS-CAPILLARY COLLIMATOR FOR DISTANCE COMPENSATION AND PARTIAL MONOCHROMATIZATION AT ROTATING-ANODE X-RAY GENERATORS

Citation
Jj. Muller et al., GLASS-CAPILLARY COLLIMATOR FOR DISTANCE COMPENSATION AND PARTIAL MONOCHROMATIZATION AT ROTATING-ANODE X-RAY GENERATORS, Journal of applied crystallography, 28, 1995, pp. 853-855
Citations number
15
Categorie Soggetti
Crystallography
ISSN journal
00218898
Volume
28
Year of publication
1995
Part
6
Pages
853 - 855
Database
ISI
SICI code
0021-8898(1995)28:<853:GCFDCA>2.0.ZU;2-8
Abstract
Access to the beam ports of rotating-anode X-ray generators is often o bstructed by direct-coupled or belt-driven target drives. The construc tion of an easily adjustable stable glass-capillary collimator is desc ribed, which renders possible the unrestricted use of beam ports of th ese generators. Transmitted intensity and monochromaticity of the prim ary beam are sufficient for precession photographs of proteins after a dditional 20 mu m Ni filtering as demonstrated by a precession photogr aph of hen egg lysozyme. The straight capillary collimator is now a ro utinely usable low-cost device for each X-ray laboratory.