Photoconductor linear array detectors for hard X-ray photon detection
were fabricated using Molecular Beam Epitaxially (MBE) grown (111)B Cd
Te on (100) Si substrates. The ohmic contact to the CdTe was excellent
and provided a linear response to the X-ray photons. The response cur
rent was linearly increased in a 7-19 keV energy range. The stability
of the detectors was very satisfactory and they were tested over an ex
tended period of time without noticeable degradation. The excellent re
sponse uniformity of a 64 linear array was obtained. The performance o
f the photoconductor was greatly improved when the detector was cooled
. At 230 K, the dynamic range of the detector response increased to ne
arly four decades.