Jpg. Piette et L. Jacques, EVALUATION OF A RAPID FAT ANALYZER, BASED ON OHMIC HEATING, FOR THE CLASSIFICATION OF GROUND-BEEF, Die Fleischwirtschaft, 76(12), 1996, pp. 1340-1342
A commercial rapid fat analyzer was evaluated for the classification o
f ground beef. The analyzer measures the volume of fat released by a g
iven amount of meat during ohmic cooking. As advertised, the equipment
was simple to operate and gave a reading in less than five minutes. W
hen used as recommended by the manufacturer, the analyzer was found to
give biased results, with lower fat contents being systematically ove
restimated while higher values were underestimated. The bias could be
partially corrected, however, by homogenizing the meat sample to a pas
te prior to the actual analysis. Thus modified, the method remained fa
st, easy to apply, and it may be useful in the routine monitoring of g
round beef, when a deviation of 2% from the actual fat value is judged
acceptable.