CONSTRUCTION OF A DOUBLE-RADIUS GUINIER DIFFRACTOMETER USING MO K-ALPHA(1) RADIATION AND A ONE-DIMENSIONAL OR 2-DIMENSIONAL DETECTION SYSTEM

Authors
Citation
Hh. Otto et W. Hofmann, CONSTRUCTION OF A DOUBLE-RADIUS GUINIER DIFFRACTOMETER USING MO K-ALPHA(1) RADIATION AND A ONE-DIMENSIONAL OR 2-DIMENSIONAL DETECTION SYSTEM, Journal of applied crystallography, 29, 1996, pp. 495-497
Citations number
6
Categorie Soggetti
Crystallography
ISSN journal
00218898
Volume
29
Year of publication
1996
Part
4
Pages
495 - 497
Database
ISI
SICI code
0021-8898(1996)29:<495:COADGD>2.0.ZU;2-S
Abstract
This paper presents the design for a computer-assisted double-radius G uinier system for Mo K alpha(1), radiation with a cylindrically bent i maging plate and an integrated readout scanner for the recorded diffra ction patterns. This quick X-ray detection technique makes new applica tions for the Guinier method possible. The addition of suitable compon ents such as an optimized monochromator, a set-up for changing, moving and heating the samples and sophisticated adjustment equipment improv es the diffractometer, making it a competitive in-house system for pow der characterization and crystal structure determination of high stand ard. The commonly used scintillation tube arrangement should also be a dapted to receive a multifunctional system.