HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY ANALYSIS OF PEANUT PHOSPHOLIPIDS .1. INJECTION SYSTEM FOR SIMULTANEOUS CONCENTRATION AND SEPARATION OF PHOSPHOLIPIDS

Citation
Ja. Singleton et Lf. Stikeleather, HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY ANALYSIS OF PEANUT PHOSPHOLIPIDS .1. INJECTION SYSTEM FOR SIMULTANEOUS CONCENTRATION AND SEPARATION OF PHOSPHOLIPIDS, Journal of the American Oil Chemists' Society, 72(4), 1995, pp. 481-483
Citations number
8
Categorie Soggetti
Food Science & Tenology","Chemistry Applied
ISSN journal
0003021X
Volume
72
Issue
4
Year of publication
1995
Pages
481 - 483
Database
ISI
SICI code
0003-021X(1995)72:4<481:HLAOPP>2.0.ZU;2-I
Abstract
This paper discusses the details of a high-performance liquid chromato graphy method for the simultaneous concentration and separation of pho spholipids or other trace compounds by direct oil injection using two different solvent systems. The system equilibrates and concentrates ph ospholipids on a silica column using hexane. At the same time, an anal ytical column is equilibrating and separating phospholipids using two binary solvent mixtures. This system eliminates a preconcentration ste p previously accomplished by solid-phase extraction, open-column chrom atography, and other previously used methodology. Other advantages inc lude: a 40% reduction in analysis time, elimination of a second transf er of labile compounds, decreased solvent use, and a simpler array of solvents to separate phospholipids. The method described has broader a pplications, such as trace organic compounds in water supplies, and tr ace metals with appropriate modifications for the particular analysis.