Rl. Ma et al., EVALUATION OF THE ANALYTICAL PERFORMANCE OF FLOW-INJECTION - FLAME ATOMIC-ABSORPTION SPECTROMETRY, Mikrochimica acta, 119(1-2), 1995, pp. 95-102
Analytical performance of flow injection - flame atomic absorption spe
ctrometry was evaluated in terms of precision and sensitivity. For dir
ect aqueous-solution introduction and alcoholic-eluate introduction af
ter on-line sorbent extraction, the influence of the interfacing param
eters was investigated, including the nebulizer uptake rate, sample-in
troduced flow rate and coupling tube length. High sensitivity could be
achieved by the exploitation of the synergetic effect of both sample
pumping and aspirating processes. A low nebulizer uptake rate combined
with a high flow rate of the introduced sample zone produced the opti
mum precision better than 2% for peak-height measurements, the optimum
rates being dependent on the nebulization characteristics of the solv
ent used. Precision and sensitivity deteriorate with increasing length
of the coupling tube.